The K 100 (W) CM Conductivity transmitter is a sophisticated single-channel instrument for measuring and control of conductive conductivity. Temperature measurement of the measuring media is possible. The instrument is available in a panel mounted housing (K100 CM) and in a wall mounted housing (K 100 W CM). Easy and safe operation by plain text menu guidance. Safety by password function. Linear and non-linear special pure water temperature compensation
Applications: Cooling And Boiler Feed Water, Waste Water Treatment, Process Water
Characteristics:
2 limit values with delay, assigned to alarm relay, 2 separate PI-controllers, Dosage check, Background illuminated two-line LC-Display for measured value and temperature, Serial interface RS 485 optional, Scaleable, glavanically isolated 0/4 .. 20 mA output, Display of relay status and error messages,Invertible alarm output, permanent or pulse contact
Technical data
Measuring parameter Conductivity (conductive)
0 .. 2.000 μS/cm, 0 .. 20.00 μS/cm, 0 .. 200.0 μS/cm,
0 .. 2.000 mS/cm, 0 .. 20.00 mS/cm, 0 .. 200.0 mS/cm
Input characteristics
Temperature measuring range -30 .. +140 °C
Temperature coefficient adjustable 0 .. 8,0%/°C or non-linear
Power supply: line voltage 24 / 115/ 230 VAC, +6/-10%, 40 .. 60 Hz, power consumption 10 VA
Output characteristics
output signal 0/4 .. 20 mA (scaleable, galvanically isolated),load max. 500 Ohm
registration range free scaleable within the measuring range serial interface RS 485 (optional)
Baud rate 9600, data format 8 bit
power supply 24 / 115/ 230 VAC, +6/-10%, 40 .. 60 Hz, power consumption 10 VA
Protection class: panel mounted housing IP 54 (front), IP 30 (housing) wall mounted housing IP 65
Ambient conditions
ambient temperature Operation 0 .. +50 °C, storage -20 .. +65°C
relative humidity max. 90% rH at 40°C (non-condensing)
protection class: panel mounted housing IP 54 (front), IP 30 (housing) wall mounted housing IP 65
Controller
control response on/off controller (adjustable hysteresis)
P/PI controller (pulse-pause, pulse-frequency or continuous output)
relays: 2 relays, each with a potential-free NO contact, max. 250V, 6 A, 550 VA
1 alarm relay with potential-free CO/NO contact, max. 250V, 6A, 550 VA
onset delay 0 .. 200 sec till controller active
digital input Controller stop via external contact
Certificates and approvals
CE-symbol Declaration of conformity: The product meets the requirements of the harmonized European standards. It thus complies with the legal requirements of the EC directives.
The manufacturer confirms successful testing of the product by affixing the CE symbol.
Mechanical construction
Panel mounted housing, material Noryl, dimensions 90x90x116 mm,
installation dimensions 92x92x140 mm, weight 0.75 kg, connection push-srew terminals
Wall mounted housing, material ABS, dimensions 165x160x85 mm, weight 0.95 kg, connection Spring-loaded terminal
Order Information
K100CM (100100K) conductive, panel mounted housing, 230 VAC
K100WCM (105100K) conductive, wall mounted housing, 230 VAC
Accessories
PKV 30-DPS Converter for PROFIBUS-DP
S-341 data logger
|
Data Sheet
Manual
|